首页 | 本学科首页   官方微博 | 高级检索  
     


A micro PDMS flow sensor based on time-of-flight measurement for conductive liquid
Authors:Haixia Yu  Dachao Li  Robert C Roberts  Kexin Xu  Norman C Tien
Affiliation:1. State Key Laboratory of Precision Measuring Technology and Instruments, Tianjin University, 300072, Tianjin, China
2. Department of Electrical Engineering and Computer Science, Case Western Reserve University, Cleveland, 44106, USA
Abstract:Transdermal extraction of interstitial fluid (ISF) offers an attractive method for non-invasive blood glucose monitoring. In order to calculate blood glucose concentration accurately, precise volume measurement of transdermally extracted ISF is required due to human skin’s varying permeability. In this paper, we presented a novel flow sensor fabricated from polydimethylsiloxane (PDMS), designed to measure the volume of conductive liquid. The flow sensor consists of two pairs of metal electrodes, which are fabricated in the PDMS channel. The volume of liquid is measured utilizing the time-of-flight of the two electrode pairs’ resistance while the liquid is flowing through the flow sensor. 1–14 μL normal saline solution was measured, the flow sensor measured volumes correlate very well (R 2 = 0.9996 and R 2 = 0.9975 for vacuum pump and syringe pump situations respectively) with the actual volumes. And the coefficient of variation for 10 times 10 μL normal saline solution measurement is 0.0077 (vacuum pump) and 0.0381 (syringe pump), respectively. The demonstrated flow sensor provides excellent functionality for conductive liquid.
Keywords:
本文献已被 SpringerLink 等数据库收录!