Comparison of the TSI Model 3306 Impactor Inlet with the Andersen Cascade Impactor: solution metered dose inhalers
Authors:Myrdal Paul B  Stein Stephen W  Mogalian Erik  Hoye William  Gupta Abhishek
Affiliation:  a College of Pharmacy, University of Arizona, Tucson, Arizona, USA b 3M Drug Delivery Systems, Early Pharmaceutics and Technology Laboratory, St. Paul, Minnesota, USA c Inhalation Formulation, Cardinal Health, Research Triangle Park, North Carolina, USA
Abstract:The product performance of a series of solution Metered Dose Inhalers (MDIs) were evaluated using the TSI Model 3306 Impactor Inlet and the Andersen Cascade Impactor (ACI). The goal of the study was to test whether the fine particle and coarse particle depositions obtained using the Model 3306 were comparable to those results obtained by ACI testing. The analysis using the Model 3306 was performed as supplied by the manufacturer as well as with 20 cm and 40 cm vertical extensions that were inserted between the Model 3306 and the USP Inlet. Nine different solution formulations were evaluated. The drug concentrations ranged from 0.08 to 0.8% w/w and the ethanol cosolvent concentration varied between 5 and 20% w/w. In general, it was found that good correlations between the two instruments were obtained. However, for formulations containing 10-20% w/w ethanol it is shown that an extension fitted to the Model 3306 yielded an improved correlation to those obtained from the ACI.
Keywords:Metered dose inhaler  MDI  Andersen cascade impactor  Impactor inlet