首页 | 本学科首页   官方微博 | 高级检索  
     


Automation of a surface sampling probe/electrospray mass spectrometry system
Authors:Kertesz Vilmos  Ford Michael J  Van Berkel Gary J
Affiliation:Organic and Biological Mass Spectrometry Group, Chemical Sciences Division, Oak Ridge National Laboratory, Tennessee 37831-6131, USA. kerteszv@ornl.gov
Abstract:An image analysis automation concept and the associated software (HandsFree TLC/MS) were developed to control the surface sampling probe-to-surface distance during operation of a surface sampling electrospray system. This automation system enables both "hands-free" formation of the liquid microjunction used to sample material from the surface and hands-free reoptimization of the microjunction thickness during a surface scan to achieve a fully automated surface sampling system. The image analysis concept and the practical implementation of the monitoring and automated adjustment of the sampling probe-to-surface distance (i.e., liquid microjunction thickness) are presented. The added capabilities for the preexisting surface sampling electrospray system afforded through this software control are illustrated by an example of automated scanning of multiple development lanes on a reversed-phase C8 TLC plate and by imaging inked lettering on a paper surface. The post data acquisition processing and data display aspects of the software package are also discussed.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号