首页 | 本学科首页   官方微博 | 高级检索  
     

基于多孔介质模型的膜式氧合器内部流场分析
引用本文:叶非华,廖虎,易国斌.基于多孔介质模型的膜式氧合器内部流场分析[J].化工进展,2020,39(3):898-905.
作者姓名:叶非华  廖虎  易国斌
作者单位:1.广东工业大学轻工化工学院,广东 广州 510006;2.广东顺德工业设计研究院(广东顺德创新设计研究院),广东 佛山 528300;3.武汉科技大学机械自动化学院,湖北 武汉 430081
基金项目:国家自然科学基金;广东省高校创新强校重大科研项目
摘    要:膜式氧合器内部流体运动特性对其性能有重要影响,利用计算流体力学(CFD)对氧合器模型进行流体动力学分析是预测其性能的重要方法之一。本文基于压降实验计算氧合器纤维束的黏性阻力系数,建立了各向同性多孔介质模型。采用RNGk-ε湍流模型对不同流量下氧合器内部流场进行计算,得到了血液速度、压力和壁面剪切应力分布云图。发现随着流量的增加,氧合器内部速度梯度分布形式基本保持不变,压力分布呈倾斜状态且逐渐减小,大部分压力损失位于纤维束内,其中53.3%位于氧合室,42.6%位于变温室。氧合器血液的入口及出口位置为血液损伤的高发区域。采用溶血数值预估模型计算得到了氧合器的标准溶血指数NIH。结果表明:在低流量1.65~3.00L/min下,各向同性多孔介质模型的模拟结果与实验结果基本一致,模拟数值与实验数值的偏差会随着液体流量的增加而变大;流量为1.65~6.00L/min时,标准溶血指数NIH为0.0084~0.0835g/100L,满足人体生理允许的使用范围。

关 键 词:  氧合器  计算流体力学  多孔介质  溶血  
收稿时间:2019-07-08

Internal flow field analysis of membrane oxygenator based on porous media model
YE Feihua,LIAO Hu,YI Guobin.Internal flow field analysis of membrane oxygenator based on porous media model[J].Chemical Industry and Engineering Progress,2020,39(3):898-905.
Authors:YE Feihua  LIAO Hu  YI Guobin
Affiliation:1.School of Chemical Engineering and Light Industry, Guangdong University of Technology, Guangzhou 510006, Guangdong, China
2.Guangdong Shunde Industrial Design Institute (Guangdong Shunde Innovative Design Institute), Foshan 528300, Guangdong, China
3.School of Machinery and Automation, Wuhan University of Science and;Technology, Wuhan 430081, Hubei, China
Abstract:The fluid motion characteristics of the membrane oxygenator have an important influence on its performance, the computational fluid dynamics (CFD) hydrodynamic analysis of the oxygenator model by computational fluid dynamics is one of the important methods to predict its performance. Based on the pressure drop experiment, the viscous drag coefficient of the oxygenator fiber bundle was calculated and an isotropic porous media model was established. The RNGk-ε turbulence model was used to calculate the internal flow field of the oxygenator under different flow rates, and the cloud velocity map of blood velocity, pressure and wall shear stress was obtained. It is found that with the increase of flow rate, the internal velocity gradient distribution of the oxygenator is basically unchanged, the pressure distribution is inclined and gradually decreases, and most of the pressure loss is located in the fiber bundle, of which 53.3% is located in the oxygenation chamber, and 42.6% is located in the temperature chamber. The inlet and outlet locations of the oxygenator blood are high-risk areas of blood damage. The standard hemolysis index NIH of the oxygenator was calculated using the hemolysis numerical prediction model. The results show that the simulation results of the isotropic porous medium model are basically consistent with the experimental results at 1.65—3.00L/min. The deviation between the simulated value and the experimental value will increase with the increase of the liquid flow rate; at 1.65—6.00L/min, the standard hemolysis index NIH is 0.0084—0.0835g/100L, which meets the physiologically acceptable use range of human body.
Keywords:membrane  oxygenator  computational fluid dynamics  porous media  hemolysis  
本文献已被 万方数据 等数据库收录!
点击此处可从《化工进展》浏览原始摘要信息
点击此处可从《化工进展》下载全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号