首页 | 本学科首页   官方微博 | 高级检索  
     


Electrical impedance tomography: regularized imaging and contrast detection
Authors:Adler A  Guardo R
Affiliation:Inst. de Genie Biomed., Ecole Polytech., Montreal, Que.
Abstract:Dynamic electrical impedance tomography (EIT) images changes in the conductivity distribution of a medium from low frequency electrical measurements made at electrodes on the medium surface. Reconstruction of the conductivity distribution is an under-determined and ill-posed problem, typically requiring either simplifying assumptions or regularization based on a priori knowledge. This paper presents a maximum a posteriori (MAP) approach to linearized image reconstruction using knowledge of the noise variance of the measurements and the covariance of the conductivity distribution. This approach has the advantage of an intuitive interpretation of the algorithm parameters as well as fast (near real time) image reconstruction. In order to compare this approach to existing algorithms, the authors develop figures of merit to measure the reconstructed image resolution, the noise amplification of the image reconstruction, and the fidelity of positioning in the image. Finally, the authors develop a communications systems approach to calculate the probability of detection of a conductivity contrast in the reconstructed image as a function of the measurement noise and the reconstruction algorithm used.
Keywords:
本文献已被 PubMed 等数据库收录!