400~1700 nm绝对光谱响应度校准和量值传递
引用本文:史学舜,徐楠,刘长明,甘海勇,李健军,林延东,陈坤峰.400~1700 nm绝对光谱响应度校准和量值传递[J].计量学报,2017,38(6):681-685.
作者姓名:史学舜  徐楠  刘长明  甘海勇  李健军  林延东  陈坤峰
作者单位:1. 中国电子科技集团公司 第四十一研究所, 山东 青岛 266555
2. 中国计量科学研究院, 北京 100029
3. 中国科学院 通用光学定标与表征技术重点实验室, 安徽 合肥 230031
基金项目:国家“863”计划(2015AA123702); 国家技术基础项目(J312013A001)
摘    要:介绍了高准确度光辐射功率校准原理和方法,利用低温辐射计作为主标准器,以陷阱探测器作为传递标准,激光器作为光源,通过激光功率稳定装置,校准了硅陷阱探测器和铟镓砷陷阱探测器的绝对光谱响应度。选取476.1, 488, 514.7, 521, 568, 632.8, 647.1, 785, 852, 980, 1064, 1550,nm共12条谱线完成了校准实验,绝对光谱响应度测量不确定度均优于0.05%。通过量子效率模型得出了硅陷阱探测器的绝对光谱响应曲线。利用InGaAs陷阱探测器分立波长点的绝对光谱响应度与相对光谱响应曲线进行了验证分析。结果表明,2种陷阱探测器均可用作传递标准进行高准确度的可见光和近红外光辐射功率校准和量值传递。

关 键 词:计量学  绝对光谱响应度  绝对低温辐射计  陷阱探测器  不确定度  
收稿时间:2015-12-25

Calibration and value transfer of absolute spectral responsivity between 400nm and 1700nm
SHI Xue-shun,XU Nan,LIU Chang-ming,GAN Hai-yong,LI Jian-jun,LIN Yan-dong,CHEN Kun-feng.Calibration and value transfer of absolute spectral responsivity between 400nm and 1700nm[J].Acta Metrologica Sinica,2017,38(6):681-685.
Authors:SHI Xue-shun  XU Nan  LIU Chang-ming  GAN Hai-yong  LI Jian-jun  LIN Yan-dong  CHEN Kun-feng
Affiliation:1. The 41st Research Institute of CETC, Qingdao, Shandong 266555, China
2. National Institute of Metrology, Beijing 100029, China
3. Key Laboratory of Optical Calibration and Characterization of CAS, Hefei, Anhui 230031, China
Abstract:The principle and methods of optical radiation measurement was illustrated. The absolute spectral responsivity of Si trap detectors and InGaAs trap detectors were calibrated with the absolute cryogenic radiometer at eleven wavelengths between 400 nm and 1700 nm, which are 476.1, 488, 514.7, 521, 568, 632.8, 647.1, 785, 852, 980, 1064 and 1550 nm, respectively, with measurement uncertainties better than 0.05%. The absolute spectral responsivity curves of Si trap detectors were calculated by quantum efficiency model. The absolute responsivity of three wavelengths was applied to get absolute spectral responsivity from the relative spectral responsivity of InGaAs detector. Results show that the two types of trap detectors can be used to calibrate and transfer optical radiation value in the visible and near infrared spectral range.
Keywords:metrology  absolute spectral responsivity  absolute cryogenic radiometer  trap detector  uncertainty  