The applicability of radiophotoluminescence dosemeter (RPLD) for measuring medical radiation (MR) doses
Authors:Manninen A-L  Koivula A  Nieminen M T
Affiliation:Physical Sciences, Biophysics, University of Oulu, Oulu, Finland. annaleem@mail.student.oulu.fi
Abstract:The applicability of radiophotoluminescence dosimetry was determined by assessing various radiophotoluminescence dosemeter (RPLD) properties for measuring medical radiation doses from radiation sources of a continuous spectrum. The RPLD was found to be accurate for measuring doses in diagnostics (50-125 keV) and radiation therapy (6, 10 and 18 MV photons, 6 and 15 MeV electrons). The RPLD shows excellent dose linearity (R(2) > 0.99), reproducibility and batch uniformity, and minimal fading and accurate accumulated dose measurement. The dosemeter material is independent of photon energy in the diagnostic range; however, the dosemeter requires additional calibration in the mammography energy range and also for accurate dose measurement with photon or electron energies in radiation therapy. RPLD measurements with a tin filter show considerable angular dependence at angles exceeding 50° between the photon beam and the normal to the long axis of the dosemeter. The RPLD measurement accuracy at high doses can be improved with optimised pre-heating schemes.