Evaluation of accuracy of measuring intraocular pressure by handheld non-contact applanation tonometer |
| |
Authors: | X Chen D Peng W Zhou Y Zhong |
| |
Affiliation: | Zhongshan Ophthalmic Center, Sun Yat-sen University of Medical Sciences, Guangzhou, China. |
| |
Abstract: | PURPOSE: To evaluate the accuracy of measuring intraocular pressure by handheld non-contact applanation tonometer. METHOD: 58 patients' (113 eyes) intraocular pressure were measured by Keeler, non-contact tonometer and R 900 Goldmann applanation tonometer and the results of measurement of intraocular pressure by the two kinds of tonometers were compared. RESULT: The mean intraocular pressure measured by non-contact is 16.31 +/- 5.59 mmHg and 17.49 +/- 6.13 mmHg (1 mmHg = 0.1333 kPa) by Goldmann applanation tonometer, respectively. There was no statistical significance to be found (P > 0.05) between the two methods. By linear correlation and regression analysis, a positive correlation was found between the two methods (r = 0.8942, b = 0.8154). CONCLUSION: The handheld non-contact tonometer has the same accuracy and reliability of measurement of intraocular pressure comparing with Goldmann applanation tonometer, and it can be used in glaucoma clinic and screening. |
| |