首页 | 本学科首页   官方微博 | 高级检索  
     


Analysis of perchlorate in human urine using ion chromatography and electrospray tandem mass spectrometry
Authors:Valentín-Blasini Liza  Mauldin Joshua P  Maple David  Blount Benjamin C
Affiliation:Division of Laboratory Sciences, National Center for Environmental Health, Centers for Disease Control and Prevention, Atlanta, Georgia 30341, USA.
Abstract:Because of health concerns surrounding widespread exposure to perchlorate, we developed a sensitive and selective method for quantifying perchlorate in human urine using ion chromatography coupled with electrospray ionization tandem mass spectrometry. Perchlorate was quantified using a stable isotope-labeled internal standard ((18)O(4)-perchlorate) with excellent assay precision (coefficient of variation <5% for repetitively analyzed quality control material). Analytical accuracy was established by blind analysis of certified proficiency testing materials prepared in synthetic urine matrix; calculated amounts deviated minimally from true amounts, with percent differences ranging from 2% to 5%. Selective chromatography and tandem mass spectrometry reduced the need for sample cleanup, resulting in a rugged and rapid method capable of routinely analyzing 75 samples/day. The lowest reportable level (0.025 ng/mL) was sufficiently sensitive to detect perchlorate in all human urine samples evaluated to date, with a linear response range from 0.025 to 100 ng/mL. This selective, sensitive, and rapid method will help elucidate any potential associations between human exposure to low levels of perchlorate and adverse health effects.
Keywords:
本文献已被 PubMed 等数据库收录!