NONLINEAR RESPONSES OF GAMMA —RAY DOSIMETERS
引用本文:罗达玲,杨健明,等.NONLINEAR RESPONSES OF GAMMA —RAY DOSIMETERS[J].核技术(英文版),1994,5(1):58-64.
作者姓名:罗达玲  杨健明
作者单位:[1]DepartmentofPhysics,ZhongshanUniversity,Guangzhou510275 [2]CitypolytechnicofHongKong,ZhongshanUniversity,Guangzhou510275
摘    要:Either sublinear or supralinear responses of dosimeters to γ-ray can be described by a response function derived from statistical Poisson distribution.The characteristic parameters of the function determine linearity,sublinearity and supralinearlty in their responses.The experimental data of gamma dose-responses of alanine ESR dosimeters film dosimeters.LiF(Mg,Cu,P) and LiF(Mg,Ti) thermoluminescence dosimeters are used to test the response function.

关 键 词:γ射线剂量计  非线性响应  剂量线性范围

NONLINEAR RESPONSES OF GAMMA-RAY DOSIMETERS
Luo Daling, E.C.M. Yong and Zhang Chunxiang.NONLINEAR RESPONSES OF GAMMA-RAY DOSIMETERS[J].Nuclear Science and Techniques,1994,5(1):58-64.
Authors:Luo Daling  ECM Yong and Zhang Chunxiang
Abstract:Either sublinear or supralinear responses of dosimeters to y-ray can be described by a response function derived from statistical Poisson distribution.The characteristic parameters of the function determine linearity, sublinearity and supralinearity in their responses. The experimental data of gamma dose-responses of alanine ESR dosimeters,film dosimeters, LiF(Mg, Cu, P) and LiF(Mg, Ti) thermoluminescence dosimeters are used to test the response function.
Keywords:Dosimeters  Nonlinear responses  y-ray