首页 | 本学科首页   官方微博 | 高级检索  
     


Polymer membrane electrodes for sensitive potentiometric determination of beta-blockers
Authors:Wassil Anwar A  Farag Abd El-Ftaah Bastawy  Moukdad Fatma A
Affiliation: a National Organization for Drug Control and Research, Cairo, Egyptb Department of Chemistry, Faculty of Science, Helowan University, Cairo, Egypt
Abstract:The construction of PVC matrix-type β-blockers (sotalol, carvedilol, and betaxolol) ion selective electrodes and their use for direct potentiometry of their respective species are described. The proposed sensors are based on the complex ion associates of β-blockers with tungstophosphate (TP) and Ammonium Reineckate (Rein) ionophoris in poly vinyl chloride membrane (PVC) with Dioctylphthalate (DOP) plasticizer. The four electrodes (Beta-TP), (Sota-TP), (Carve-TP), and (Cave-Rein) show stable potential response with near Nernstian slope of 50.8, 33.7, 32.35, and 33 mv per decade, range of concentration 10-2-10-7 M β-blockers. Selectivity coefficients data obtained for 11 different organic and inorganic ions are presented. The electrodes have fast response time (30 and 40 s) and were used over wide range of pH 4.5-8.5. Validation of the method according to the quality assurance standers shows suitability of proposed sensors for use in the quality control assessment of these drugs. The results obtained for the determination of β-blockers with the proposed electrodes show average recoveries of 100.78% and a mean standard deviation of ±1.2. The nominal are obtained. The data agree well with those obtained by standard methods.
Keywords:Determination of β-blockers  Potentiometry  Tungstophosphate  Reineckate  Pharmaceutical analysis
本文献已被 InformaWorld PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号