首页 | 本学科首页   官方微博 | 高级检索  
     


On the determination of an optimum microwave diathermy frequency for a direct contact applicator
Authors:A W Guy  J F Lehmann
Affiliation:Department of Physical Medicine and Rehabilitation, School of Medicine, University of Washington, Seattle, Wash.;
Abstract:An approximate theoretical analysis is presented for determining the relative heating in a two layer fat-muscle medium due to a dielectric loaded dipole-corner reflector applicator in direct contact with the finite fat layer. The results indicate that convenient applicator sizes, equivalent to those now used at 2450 Mc/s, operating at a frequency in the neighborhood of 750 Mc/s provide the maximum ratio of relative heat per unit volume in the muscle as compared to that in the fat. Experimental data taken with the aid of an electrical model of the fat-muscle layers shows reasonable agreement with the results of the approximate analysis.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号