首页 | 本学科首页   官方微博 | 高级检索  
     


Microchip for combining gas chromatography or capillary liquid chromatography with atmospheric pressure photoionization-mass spectrometry
Authors:Haapala Markus  Luosujärvi Laura  Saarela Ville  Kotiaho Tapio  Ketola Raimo A  Franssila Sami  Kostiainen Risto
Affiliation:Division of Pharmaceutical Chemistry, Faculty of Pharmacy, P.O. Box 56, FI-00014 University of Helsinki, Finland.
Abstract:We present a microfabricated nebulizer chip for combining atmospheric pressure photoionization-mass spectrometry (APPI-MS) with gas chromatography (GC) or capillary liquid chromatography (capLC). The chip consists of a silicon plate and a glass plate or two glass plates. The chip includes a sample inlet channel, auxiliary gas and dopant inlet, vaporizer channel, nozzle, and platinum heater. The sample eluted from the capLC or GC is mixed with auxiliary gas and dopant (toluene) in the heated vaporizer. The chip forms a confined jet of the sample vapor, which is photoionized as it exits the chip. The analytical performance of GC- and capLC-microchip APPI-MS was evaluated with some polycyclic aromatic hydrocarbons, amphetamines, and steroids. The GC-muAPPI-MS method provides high sensitivity down to 0.8 fmol, repeatability (RSD = 7.5-14%), and linearity (r = 0.9952-0.9987). The capLC-muAPPI-MS method shows high sensitivity down to 1 fmol, good repeatability (RSD = 3.6-8.1%), and linearity (r = 0.9989-0.9992).
Keywords:
本文献已被 PubMed 等数据库收录!