首页 | 本学科首页   官方微博 | 高级检索  
     


Simultaneous concentration and separation of coumarins using a molecular micelle in micellar affinity gradient focusing
Authors:Kamande Mary W  Ross David  Locascio Laurie E  Lowry Mark  Warner Isiah M
Affiliation:Department of Chemistry, Louisiana State University, Baton Rouge, Louisiana 70803, USA.
Abstract:We report the use of a molecular micelle for the simultaneous separation and concentration of neutral and hydrophobic analytes using micellar affinity gradient focusing (MAGF). The technique, MAGF, combines the favorable features of micellar electrokinetic chromatography and temperature gradient focusing. The focusing of neutral coumarin analytes was accomplished by the use the molecular micelle, poly(sodium undecenyl sulfate) (poly-SUS). Concentration enhancements of 10-25-fold/min were achieved for focusing of the coumarin dyes. The effect of varying the temperature gradient on the resolution of two of the coumarin dyes was also investigated, demonstrating that improved resolution could be achieved by reducing the steepness of the temperature gradient. In addition, with scanning-mode MAGF (in which the peaks are sequentially scanned past a fixed detection point by varying the buffer counterflow velocity), the use of poly-SUS was shown to produce repeatable and quantitative analyte peaks, making quantitative separations possible with the MAGF technique. Finally, it was shown that peak areas could be increased in scanning MAGF by reducing the scan rate so that the sensitivity of the method can be adjusted as needed.
Keywords: