首页 | 本学科首页   官方微博 | 高级检索  
     


Experimental comparison of different dead-time correction techniques in single-channel counting experiments
Authors:C. Michotte  M. Nonis
Affiliation:aBureau International des Poids et Mesures, Pavillon de Breteuil, 92312 Sèvres Cedex, France
Abstract:For single-channel counting, the dead-time correction method using transmission formulae was compared experimentally with the live-time technique. Agreement is within 4×10−4 for count rates up to 104 s−1 when the formulae for dead times in series are used and the dead-time value related to the analogue pulse width is adjusted for the BIPM modules. At higher count rates, a bias of 2×10−3 is observed. Two different live-time modules were compared and it was demonstrated that the live-time correction is also sensitive to the analogue pulse width and to the duration or rise time of the clock pulses. When these effects are taken into account, the two modules differ by only 5×10−4.
Keywords:(Non-)extendable dead time   Dead times in series   Dead-time transmission formulae   Live-time correction   Analogue pulse width
本文献已被 ScienceDirect 等数据库收录!