Four-point probe characterization of 4H silicon carbide |
| |
Authors: | N. Chandra V. SharmaG.Y. Chung D.K. Schroder |
| |
Affiliation: | a School of Electrical, Computer and Energy Engineering, Arizona State University, Tempe, AZ 85287-5706, USA b Dow Corning Corp., Midland, MI 48611, USA |
| |
Abstract: | We report on four-point probe measurements on SiC wafers as such measurements give erratic data. Current-voltage measurements on n-type SiC wafers doped to 3 × 1018 cm−3 are non-linear and single probe I-V measurements are symmetrical for positive and negative voltages. For comparison, similar measurements of p-type Si doped to 5 × 1014 cm−3 gave linear I-V, well-defined sheet resistance and the single probe I-V curves were asymmetrical indicating typical Schottky diode behavior. We believe that the reason for the non-linearity in four-point probe measurements on SiC is the high contact resistance. Calculations predict the contact resistance of SiC to be approximately 1012 Ω which is of the order of the input resistance of the voltmeter in our four-point probe measurements. There was almost no change in two-probe I-V curves when the spacing between the probes was changed from 1 mm to 2 cm, further supporting the idea that the I-V characteristics are dominated by the contact resistance. |
| |
Keywords: | Silicon carbide Four-point probe Thermionic-field emission Contact resistance |