CT dosimetry computer codes: their influence on radiation dose estimates and the necessity for their revision under new ICRP radiation protection standards
Authors:Kim Kwang Pyo  Lee Jaekook  Bolch Wesley E
Affiliation:Department of Nuclear Engineering, Kyung Hee University, Seocheon-dong, Giheung-gu, Yongin-si, Gyeonggi-do, Korea. kpkim@khu.ac.kr
Abstract:Computed tomography (CT) dosimetry computer codes have been most commonly used due to their user friendliness, but with little consideration for potential uncertainty in estimated organ dose and their underlying limitations. Generally, radiation doses calculated with different CT dosimetry computer codes were comparable, although relatively large differences were observed for some specific organs or tissues. The largest difference in radiation doses calculated using different computer codes was observed for Siemens Sensation CT scanners. Radiation doses varied with patient age and sex. Younger patients and adult females receive a higher radiation dose in general than adult males for the same CT technique factors. There are a number of limitations of current CT dosimetry computer codes. These include unrealistic modelling of the human anatomy, a limited number of organs and tissues for dose calculation, inability to alter patient height and weight, and non-applicability to new CT technologies. Therefore, further studies are needed to overcome these limitations and to improve CT dosimetry.
Keywords:
本文献已被 PubMed 等数据库收录!