首页 | 本学科首页   官方微博 | 高级检索  
     


An accurate technique of measurement of a transducer output by using a modified two core saturable reactor
Authors:SC Bera  N Mandal  R Sarkar  
Affiliation:aInstrumentation Engineering Section, Department of Applied Physics, University of Calcutta, 92 A.P.C. Road, Kolkata 700 009, India;bAsansol Engineering College, Vivekananda Road, Kanyapur, Asansol 713 305, West Bengal, India
Abstract:Saturable reactor or transductor used as a magnetic modulator or amplifier for the measurement of the small value DC output signal of a transducer may suffer from error due to the fact that a part of the current drawn from the transducer by the transductor unit is utilized to magnetize the core material. In this paper a modified two core saturable reactor circuit has been proposed to eliminate this error. This unit also minimizes some other errors like common mode error, material error, etc. Theory of operation and design of the circuit has been explained in this paper. The circuit has been tested by using a millivolt input signal as well as a thermocouple to measure temperature. The experimental results are presented in this paper. Linear static characteristic graphs have been observed.
Keywords:Saturable reactor  Feedback technique  Transducer  Small DC signal measurement  Thermocouple
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号