首页 | 本学科首页   官方微博 | 高级检索  
     


The monopole-source solution for estimating tissue temperatureincreases for focused ultrasound fields
Authors:Ellis   D.S. O'Brien   W.D.   Jr.
Affiliation:Dept. of Electr. & Comput. Eng., Illinois Univ., Urbana, IL;
Abstract:The monopole-source solution to the problem of estimating tissue temperature rise generated by a focused ultrasound beam is presented. The acoustic pressure field generated by a focused, continuous-wave ultrasound source using the acoustic monopole-source method is developed. The point-source solution to the linear bio-heat transfer equation is used to calculate the axial, steady-state temperature increase for both circular and rectangular apertures. The results of the circular aperture are compared with the temperature increase calculated using the heated-disc method and are shown to be in substantial agreement. Finally, the temperature increase generated by the circular aperture is compared to that of the rectangular aperture for the same source power, aperture surface area, operating frequency, and medium properties, and it is shown that the rectangular source generates temperature increases less than those of the circular source under these conditions
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号