首页 | 本学科首页   官方微博 | 高级检索  
     


General and self-consistent method for the calibration of polarization modulators, polarimeters, and mueller-matrix ellipsometers
Authors:Compain E  Poirier S  Drevillon B
Affiliation:Laboratoire de Physique des Interfaces et des Couches Minces, Unite Mixte de Recherche du Centre National de la Recherche Scientifique, Ecole Polytechnique, 91128 Palaiseau, France.
Abstract:Calibration of polarization-state generators (PSG's), polarimeters, and Mueller-matrix ellipsometers (MME's) is an important factor in the practical use of these instruments. A new general procedure, the eigenvalue calibration method (ECM), is presented. It can calibrate any complete MME consisting of a PSG and a polarimeter that generate and measure, respectively, all the states of polarization of light. In the ECM, the PSG and the polarimeter are described by two 4 x 4 matrices W and A, and their 32 coefficients are determined from three or four measurements performed on reference samples. Those references are smooth isotropic samples and perfect linear polarizers. Their optical characteristics are unambiguously determined during the calibration from the eigenvalues of the measured matrices. The ECM does not require accurate alignment of the various optical elements and does not involve any first-order approximation. The ECM also displays an efficient error control capability that can be used to improve the MME behavior. The ECM is illustrated by an experimental calibration, at two wavelengths (458 and 633 nm), of a MME consisting of a coupled phase modulator associated with a prism division-of-amplitude polarimeter.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号