Detection of copper in water using on-line plasma-excited atomic absorption spectroscopy (AAS) |
| |
Authors: | Porento Mika Sutinen Veijo Julku Timo Oikari Risto |
| |
Affiliation: | University of Oulu, Measurement and Sensor Laboratory, Technology Park, Kajaani, Finland. |
| |
Abstract: | A measurement method and apparatus was developed to measure continuously toxic metal compounds in industrial water samples. The method was demonstrated by using copper as a sample metal. Water was injected into the sample line and subsequently into a nitrogen plasma jet, in which the samples comprising the metal compound dissolved in water were decomposed. The transmitted monochromatic light was detected and the absorbance caused by copper atoms was measured. The absorbance and metal concentration were used to calculate sensitivity and detection limits for the studied metal. The sensitivity, limit of detection, and quantification for copper were 0.45 ± 0.02, 0.25 ± 0.01, and 0.85 ± 0.04 ppm, respectively. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|