首页 | 本学科首页   官方微博 | 高级检索  
     


Small Multiple Fixed-Point Cell as Calibration Reference for a Dry Block Calibrator
Authors:S Marin  M Hohmann  T Fröhlich
Affiliation:1.Institut für Prozessmess- und Sensortechnik,Technische Universit?t Ilmenau,Ilmenau,Germany
Abstract:A small multiple fixed-point cell (SMFPC) was designed to be used as in situ calibration reference of the internal temperature sensor of a dry block calibrator, which would allow its traceable calibration to the International Temperature Scale of 1990 (ITS-90) in the operating range of the block calibrator from \(70\,^{\circ }\hbox {C}\) to \(430\,^{\circ }\hbox {C}\). The ITS-90 knows in this temperature range, three fixed-point materials (FPM) indium, tin and zinc, with their respective fixed-point temperatures (\(\vartheta _\mathrm {FP}\)), In (\(\vartheta _\mathrm {FP}\,{=}\,156.5985\,^{\circ }\hbox {C}\)), Sn (\(\vartheta _\mathrm {FP}\,{=}\,231.928\,^{\circ }\hbox {C}\)) and Zn (\(\vartheta _\mathrm {FP}\,{=}\,419.527\,^{\circ }\hbox {C}\)). All of these FPM are contained in the SMFPC in a separate chamber, respectively. This paper shows the result of temperature measurements carried out in the cell within a period of 16 months. The test setup used here has thermal properties similar to the dry block calibrator. The aim was to verify the metrological properties and functionality of the SMFPC for the proposed application.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号