Development of controlled Nd:YAG laser for medical applications |
| |
Authors: | S R Mordon A H Cornil B Buys J P Sozanski J M Brunetaud Y Moschetto |
| |
Affiliation: | INSERM U 279, Lille, France. |
| |
Abstract: | Several medical fields are concerned with applications of thermal lasers such as neodymium-doped, yttrium aluminum garnet (Nd:YAG), argon, and CO2. However, quantification of the necrotic volume of Nd:YAG laser-induced damage is not possible at the time of treatment. Mathematic models and feedback control can help to optimize Nd:YAG laser treatments. We therefore formulated mathematic models for coagulation processes and developed an intelligent Nd:YAG laser system with closed-loop feedback control. Surface temperature evolution proved to be valuable data for real-time control of coagulation and ablation. Infrared thermometry provided the noncontact measurement of temperature. A computer stored the temperature data calculated by the mathematic model. Deviations of surface temperature during the treatment beyond established tolerances causes the Nd:YAG laser system to adjust the laser power automatically. |
| |
Keywords: | |
|
|