A self-calibrating radon monitor with statistical discrimination |
| |
Authors: | N Valcov and L Purghel |
| |
Affiliation: | Institute of Atomic Physics, P.O. Box MG-6, R-76900 Bucharest, Magurele, Romania |
| |
Abstract: | A radon monitor, able to perform the measurement of the radon and its progeny volumic activity, in a gamma-ray or natural radiation background field, was developed. The instrument consists of a 10 l ionization chamber, a high voltage source, an integrating preamplifier, a data acquisition system and a personal computer. A new method for self-calibration of Radon volumic activity measurements, based on the alpha counting with an ionization chamber is also presented. |
| |
Keywords: | Radon Ionization chambers Statistical discrimination |