首页 | 本学科首页   官方微博 | 高级检索  
     


Determination of haloperidol drug in ampoules and in urine samples using a potentiometric modified carbon paste electrode
Affiliation:1. Chemistry Department, Faculty of Science, Cairo University, 12613 Giza, Egypt;2. Central Laboratory for Evaluation of Veterinary Biologics, Abbassia, 131, Cairo, Egypt;3. Environmental Monitoring Laboratory, Ministry of Health, Port Said, Egypt
Abstract:A modified carbon paste electrode for haloperidol drug based on haloperidol-phosphomolybdate (HP-PM) as an ion-exchanger dissolved in plasticizer DBP and its potentiometric characteristics were discussed. The electrode exhibited a good Nernstian slope of 56.9 ± 0.3 mV/decade with a linear concentration range from 3.2 × 10?6 to 1.0 × 10?2 M for the haloperidol ion. The limit of detection (LOD) was 1.5 × 10?6 M. It had response time of 5–8 seconds (s), useable in pH range of 6.2–8.6 and temperature of 20–60 °C. The electrode shows clear discrimination of haloperidol drug from several inorganic ions, sugars and some common drug excipients. The sensor was applied for determination of haloperidol drug in urine and in pharmaceutical formulations using potentiometric determination, standard addition and the calibration curve methods. The results are satisfactory with excellent percentage recovery comparable or better than those obtained by other routine methods.
Keywords:Potentiometry  Haloperidol drug  Carbon paste electrode  Ion-selective electrode
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号