Estimation of Electron-loss and Photon-scattering Corrections for Parallel-plate Free-air Chambers
Abstract:Parallel-plate free-air ionization chambers are used for X-ray air-kerma rate standards at the National Metrology Institute of Japan (NMIJ), AIST. The electron-loss and photon-scattering correction factors are needed for the evaluation of air-kerma rate from measured current. The electron-loss correction factor (Ke) is a correction of the charge loss by giving energy to the electrode part without the high-speed electron stopping in the air area where the charges are collected. The scattering correction factor (Ksc) is for a correction of extra charges produced by scattered photons generated after an incidence photon is interactive. The electron-loss and photon-scattering correction factors for 3 different size parallel-plate free-air chambers are estimated by the EGS4 code. One chamber is used to get primary standards for absolute measurements of air kerma in beams of medium-energy X-rays and the other two are used for the same purpose in beams of low-energy X-rays. These correction factors are calculated for mono-energetic photons. It is found that electron-loss and photon-scattering correction factors depend on the chamber size, and the latter especially changes greatly depending on the size. The Ke and Ksc value for medium-and low-energy X-ray fields at AIST are estimated by averaging the energy deposition contributions over the X-ray spectrum.
Keywords:free-air chamber  EGS4, Monte-Carlo simulation  electron-loss  photon-scattering  correction factor