首页 | 本学科首页   官方微博 | 高级检索  
     


In Vivo and In Vitro Degradation Behavior of Magnesium Alloys as Biomaterials
Authors:Dingchuan Xue  Yeoheung Yun  Zongqing Tan  Zhongyun Dong  and Mark JSchulz
Affiliation:1) Department of Chemical and Materials Engineering,University of Cincinnati,OH 45221,USA 2) Department of Bioengineering,North Carolina A & T State University,Greensboro,NC 27411,USA 3) Department of Internal Medicine,University of Cincinnati,OH 45221,USA 4) Department of Mechanical Engineering,University of Cincinnati,OH 45221,USA
Abstract:The corrosion behavior of pure Mg,AZ31,and AZ91D were evaluated in various in vitro and in vivo environments to investigate the potential application of these metals as biodegradable implant materials.DC polarization tests and immersion tests were performed in different simulated body solutions,such as distilled(DI) water,simulated body fluid(SBF) and phosphate buffered solution(PBS).Mg/Mg alloys were also implanted in different places in a mouse for in vivo weight loss and biocompatibility investigations.The in vivo subcutis bio-corrosion rate was lower than the corrosion rate for all of the in vitro simulated corrosive environments.The Mg/Mg alloys were biocompatible based on histology results for the liver,heart,kidney,skin and lung of the mouse during the two months implantation.Optical microscopy and scanning electron microscopy were carried out to investigate the morphology and topography of Mg/Mg alloys after immersion testing and implantation to understand the corrosion mechanisms.
Keywords:Mg/Mg alloys  In vitro/in vivo testing  Bio-degradation rate  Biocompatibility
本文献已被 CNKI ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏