首页 | 本学科首页   官方微博 | 高级检索  
     


New data on direct ion storage dosemeters
Authors:Fuchs Hermann  Stähler Simon  Dittmar Michael
Affiliation:TU Wien, Haberlgasse 77/10, A-1160 Vienna, Austria. hermann.fuchs@student.tuwien.ac.at
Abstract:The DIS-1 dosemeter from the Finnish company RADOS is an innovative kind of passive electronic dosemeter for photon and beta radiation. This study examines the 'long-term' linear response behaviour, the calibration and readout accuracy with large samples of 'used' DIS-1 dosemeters especially in the low-dose region, which is of special interest for radiation protection issues. Our measurements prove the adequacy of the DIS-1 dosemeter for long-term-personal dosimetry. The fast and precise readout seems to make the DIS-1 dosemeter an ideal choice for personal dosimetry in low-dose environments.
Keywords:
本文献已被 PubMed Oxford 等数据库收录!