首页 | 本学科首页   官方微博 | 高级检索  
     


Two-color CO2/HeNe laser interferometer for C-2 experiment
Authors:Gornostaeva O  Deng B H  Garate E  Gota H  Kinley J  Schroeder J  Tuszewski M
Affiliation:Tri Alpha Energy, Inc., P.O. Box 7010, Rancho Santa Margarita, California 92688, USA.
Abstract:A six-channel two-color interferometer has been developed for plasma electron density measurements in the C-2 field reversed configuration experiment. A CO(2) laser is utilized as the main probe beams, while copropagating visible HeNe laser beams are mainly sensitive to vibration. Density measurements in C-2 plasmas have shown that this is a reliable turn-key system. The maximum residual phase noise after vibration compensation is less than ±5°, corresponding to a line integral density of 3×10(18)?m(-2). The time resolution for routine operation is 2?μs.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏