首页 | 本学科首页   官方微博 | 高级检索  
     


Grating spectrometer system for beam emission spectroscopy diagnostics using high-energy negative-ion-based neutral beam injection on LHD
Authors:Kado S  Oishi T  Yoshinuma M  Ida K
Affiliation:School of Engineering, The University of Tokyo, 7-3-1 Hongo, Bunkyo-ku, Tokyo 113-8656, Japan. kado@n.t.u-tokyo.ac.jp
Abstract:A beam emission spectroscopy (BES) system was developed for density gradient and fluctuation diagnostics in the Large Helical Device (LHD). In order to cover the large Doppler shift of the Hα beam emission because of the high-energy negative-ion-based neutral beam atom (acceleration voltage V(acc)=90-170?kV) and the large motional Stark splitting due to the large v×B field (magnetic field B=3.0?T), a grating spectrometer was used instead of a conventional interference filter system. The reciprocal linear dispersion is about 2 nm/mm, which is sufficient to cover the motional Stark effect spectra using an optical fiber with a diameter of 1 mm.
Keywords: