首页 | 本学科首页   官方微博 | 高级检索  
     


Impurity temperature correction factors for the transmission grating spectrometer in the TJ-II stellarator
Authors:Arévalo J  McCarthy K J  Carmona J M  Fontdecaba J M
Affiliation:Laboratorio Nacional de Fusio?n, Association Euratom-Ciemat, E-28040 Madrid, Spain. juan.arevalo@ciemat.es
Abstract:Impurity ion temperature and velocity profiles are obtained across plasmas in the TJ-II stellarator by performing charge-exchange recombination spectroscopy with a diagnostic neutral beam injector. For this, a tridirectional (toroidal plus two poloidal opposing views) multichannel spectroscopic diagnostic, incorporating 12-way fiber arrays, a compact f/1.8 spectrograph, and a back-illuminated CCD, permits Doppler line shifts and widths (of the C VI line at 529.05 nm) to be determined with 1-2 cm spatial resolution. For good photon counting statistics under Li-coated wall conditions, 600?μm diameter fibers collect and transmit light to curved 100?μm wide input slits. When calibrated with a neon pencil lamp this entrance slit width results in a non-Gaussian instrumental function that, if not handled correctly, can result in systematically underestimated impurity temperatures. Here we develop and present correction factors for this effect for a range of conditions.