首页 | 本学科首页   官方微博 | 高级检索  
     


Theoretical and practical performance of a concentric bifocal intraocular implant lens
Authors:WN Charman  IJ Murray  M Nacer  EP O'Donoghue
Affiliation:Department of Optometry and Vision Sciences, UMIST, Manchester, UK. mjcijm@fs1.op.umist.ac.uk
Abstract:Some results from a survey of 43 patients who had a monofocal intra-ocular-lens (IOL) in one eye and a concentric bifocal intra-ocular-lens in the fellow eye are reported. Twenty patients with 6/9 or better post-operative visual acuity in both eyes, participated in the main part of the study. Optical transfer functions for the bifocal lens showed that, compared to an optimal single-vision correction, there is a 50% contrast degradation of the distance retinal image across all spatial frequencies above around 3 c/deg. For the patients in the main study, there was a close correspondence between practical measurement of contrast sensitivity and the theoretical predictions of the modulation transfer functions. Measuring contrast sensitivity proved an effective means of assessing misalignment of the bifocal IOLs.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号