首页 | 本学科首页   官方微博 | 高级检索  
     


Voltage ratio technique used with the national RF standardattenuation measuring equipment at the PTB
Authors:Ruhle  W Stumpe  D
Affiliation:Physikalisch Tech. Bundesanstalt, Braunschweig ;
Abstract:With a new measuring system, the attenuation range of the national RF standard attenuation measuring equipment at the Physikalisch-Technische Bundesanstalt (PTB), Germany, has been extended to 60 db in the waveguide bands R 220 and R 320. Most of the system is automated to shorten the calibration and measurement time. The total uncertainty of measurement is estimated to be 0.0027 dB per 10 dB attenuation
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号