Measurement and Simulation of TEPC Microdosimetric Spectra in Gamma Radiation Field |
| |
Authors: | ZHANG Wei-hua WANG Zhi-qiang LIU Yi-na XIAO Xue-fu LI Chun-juan LUO Hai-long |
| |
Affiliation: | China Institute of Atomic Energy, Beijing 102413, China |
| |
Abstract: | According to the different characteristics of microdosimetric spectra measured by tissue equivalent proportional counter (TEPC), the neutron dose equivalent and γ dose equivalent could be distinguished in a unknown neutron and γ mixed radiation field. In order to discriminate the γ radiation dose equivalent from the total value,the pure γ microdosimetric spectra was measured in 60Co、137Cs radionuclide radiation field with TEPC. TEPC microdosimetric spectra in a series of monoenergy γ radiation field were simulated by FLUKA code. All the γ radiation microdosimetric spectra, including measured spectrum in 60Co、137Cs radiation field and that of simulation spectrum by FLUKA code, reveal a trait that the linear energy of γ radiation is basically lower than 10 keV/μm. This trait is the very foundation to discriminate the γ radiation from the mixed radiation. |
| |
Keywords: | TEPC microdosimetric spectra &gamma radiation FLUKA code |
|
| 点击此处可从《同位素》浏览原始摘要信息 |
|