首页 | 本学科首页   官方微博 | 高级检索  
     


Technical Note: On the Calibration of an Opticle Particle Counter
Authors:Norman F Robinson  Dennis Lamb
Affiliation:Desert Research Institute, Atmospheric Sciences Center , P. O. Box 60220, Reno , NV , 89506
Abstract:Mie scattering theory is used to calculate the relative optical responses for latex and water particles in the diameter range 0.3 to 15 μm, using geometry applicable to a Royco 225 Optical Particle Counter. The results are presented in a way that permits direct correction of the instrumental calibration curve for the refractive index differences associated with the calibration and sample aerosol particles
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号