使用单能X射线辐射装置对CdTe探测效率的实验刻度
引用本文:余涛,郭思明,周建斌,蒋政,郄晓雨,郭锴悦,吴金杰. 使用单能X射线辐射装置对CdTe探测效率的实验刻度[J]. 计量学报, 2022, 43(10): 1366-1370. DOI: 10.3969/j.issn.1000-1158.2022.10.18
作者姓名:余涛  郭思明  周建斌  蒋政  郄晓雨  郭锴悦  吴金杰
作者单位:成都理工大学,四川 成都610059;中国计量科学研究院,北京100029;中国计量科学研究院,北京100029;成都理工大学,四川 成都610059;中国计量科学研究院,北京100029;河北科技大学,河北 石家庄050018;中国计量科学研究院,北京100029;中国计量大学,浙江 杭州310018
基金项目:中国计量科学研究院基本科研业务费(AKYZZ2115)
摘    要:CdTe探测器对单能平行光子源的绝对测量之前,需要进行效率刻度。利用MCNP5蒙特卡罗模拟程序建立CdTe探测器物理模型,模拟计算了10~260keV能量段能点的本征探测效率,在10~60keV能量段探测效率高于75%。用单能X射线装置和HPGe探测器对CdTe探测器本征探测效率进行了实验刻度。结果表明,在10~100keV能量范围内CdTe探测器的模拟效率与实验效率趋势一致,最大误差不超过5.6%。因为Te元素在27keV和32keV处会产生逃逸峰,导致探测效率在这2个能量处有明显下降趋势。用241Am和133Ba放射源对CdTe探测器进行效率刻度验证,在能量为59.54keV和81keV放射源标定的探测效率与单能X射线辐射装置测量值相符。

关 键 词:计量学  CdTe探测器  本征探测效率  蒙特卡罗  探测效率刻度
收稿时间:2022-03-07

Experimental Calibration of CdTe Detection Efficiency Using Single-energy X-ray Radiation Device
YU Tao,GUO Si-ming,ZHOU Jiang-bin,JIANG Zheng,QIE Xiao-yu,GUO Kai-yue,WU Jing-jie. Experimental Calibration of CdTe Detection Efficiency Using Single-energy X-ray Radiation Device[J]. Acta Metrologica Sinica, 2022, 43(10): 1366-1370. DOI: 10.3969/j.issn.1000-1158.2022.10.18
Authors:YU Tao  GUO Si-ming  ZHOU Jiang-bin  JIANG Zheng  QIE Xiao-yu  GUO Kai-yue  WU Jing-jie
Affiliation:1. Chengdu University of Technology, Chengdu, Sichuan 610059, China
2. National Institute of Metrology, Beijing 100029, China
3. Hebei University of Science and Technology, Shijiazhuang, Hebei 050018, China
4. China Jiliang University, Hangzhou, Zhejiang 310018, China
Abstract:Before realizing the absolute measurement of the single-energy parallel photon source, it is necessary to calibrate the efficiency of the CdTe detector. The physical model of the CdTe detector was established by using the MCNP5 Monte Carlo simulation program, the intrinsic detection efficiency of the energy point in the 10~260keV energy range was simulated and calculated, the detection efficiency was higher than 75% in the energy range of 10 ~ 60keV. The intrinsic detection efficiency of the CdTe detector was experimentally calibrated with a single energy X-ray device and a HPGe detector. The results show that the simulated efficiency of the CdTe detector is consistent with the experimental efficiency trend in the energy range of 10 ~ 100keV, and the maximum error does not exceed 5.6%. Because the Te element produces escape peaks at 27keV and 32keV, the detection efficiency has a clear downward trend at these two energies. Efficiency calibration of CdTe detector was carried out with 241Am and 133Ba radioactive sources, and the calibration efficiency of radioactive sources with energies of 59.54keV and 81keV was consistent with the measured value of single energy X-ray radiation device.
Keywords:metrology  CdTe detector  intrinsic detection efficiency  monte carlo  detection efficiency scale  