商品化酶抑制-比色法农残速测试剂盒对农药的敏感性测定
引用本文:罗俊霞,赵建波,唐国林,辛焕发,宋晓爽,张刚,梁爽.商品化酶抑制-比色法农残速测试剂盒对农药的敏感性测定[J].化学试剂,2021,43(1):83-88.
作者姓名:罗俊霞  赵建波  唐国林  辛焕发  宋晓爽  张刚  梁爽
作者单位:郑州市农产品质量检测流通中心,河南郑州450006;郑州市农业技术推广中心,河南郑州450006;深圳市安鑫宝科技发展有限公司,广东深圳518109;河南恒晟检测技术有限公司,河南郑州451100
基金项目:郑州市农产品质量检测中心建设项目(豫发改农经[2011]1778号)。
摘    要:探讨商品化的酶抑制-比色法农残速测试剂盒对50种农药的敏感性测定.使用商品化的酶抑制.比色法试剂盒对嘧啶磷等40种有机磷和10种氨基甲酸酯类农药残留进行5个浓度水平的测定;对有3个浓度水平的抑制率大于0的农药,用抑制率和农药浓度的常用对数进行线性拟合,得出相关系数和抑制率方程,并据此计算IC10和IC50,通过IC10...

关 键 词:酶抑制-比色法  农药残留  重现性  灵敏度  复检

Determination of Pesticide Sensitivity on Commercial Enzyme Inhibition Colorimetric Pesticide Residue Test Kit
LUO Jun-xia,ZHAO Jian-bo,TANG Guo-lin,XN Huan-fa,SONG Xiao-shuang,ZHANG Gang,LIANG Shuang.Determination of Pesticide Sensitivity on Commercial Enzyme Inhibition Colorimetric Pesticide Residue Test Kit[J].Chemical Reagents,2021,43(1):83-88.
Authors:LUO Jun-xia  ZHAO Jian-bo  TANG Guo-lin  XN Huan-fa  SONG Xiao-shuang  ZHANG Gang  LIANG Shuang
Affiliation:(Zhengzhou Agricultural Product Quality Testing and Distribution Center,Zhengzhou 450006,China;Zhengzhou Agricultural Technology Extension Centre,Henan 450006,China;Shenzhen Anxinbao Technology Development Co.,Ltd.,Shenzhen 518109,China;Henan Hengsheng Testing Technology Co.,Ltd.,Zhengzhou 451100,China)
Abstract:To investigate the sensitivity of commercial enzyme inhibitor-colorimetry agrochemical residue test kit to 50 kinds of pesticides.It was tested with commercial enzyme inhibition colorimetric kit at five concentration levels in 40 kinds of organophosphorus and 10 kinds of carbamate pesticide residues,such as pyrimidine phosphate.For pesticides with inhibition rates greater than 0 at three concentration levels,the common logarithm of inhibition rate and pesticide concentration was used for linear fitting to obtain the correlation coefficient and inhibition rate equation,and then IC10 and IC50 were calculated accordingly,and the sensitivity of the kit to various pesticides was compared by IC10.The selected kits have different sensitivities to each pesticide.For IC10,there is also a difference between 3 mL and 2.5 mL sample solutions,and the difference of individual pesticides is one or two orders of magnitude.It is recommended that users test the kit as needed when purchasing the kit.
Keywords:enzyme inhibition colorimetry  pesticide residue  reproducibility  sensitivity  retest