Target preparation for milligram sized 14C samples and data evaluation for AMS measurements
Authors:Michael Andree  Jürg Beer  Hans Oeschger  G. Bonani  H.J. Hofmann  E. Morenzoni  M. Nessi  M. Suter  W. Wölfli
Affiliation:Physics Institute of Berne, CH-3012 Berne, Switzerland;Institut für Mittelenergiephysik, ETH—Hönggerberg, CH-8093 Zürich, Switzerland
Abstract:Our preparation technique produces in a glow-discharge an amorphous carbon deposit on a copper substrate. The process starts with 1.6 cm3 CO2 STP (900 μg carbon) which is reduced over hot zinc to CO and subsequently cracked in the discharge. The yield of the process is typically 80%. With these targets in the Zürich ion source ion currents up to 20 μA are obtained. The background of samples prepared with this technique is presently around 30 ka (2.5% MODERN). The precision after half an hour measuring time for a modern sample is 0.7% and 2.7% for a three half-lives old sample, including the errors of the background and the NBS oxalic acid measurement. The method we use to correct for the background of the preparation and the accelerator as well as for the fractionation in the accelerator is presented.