首页 | 本学科首页   官方微博 | 高级检索  
     


Lower frequency of free-field calibration of hydrophones in emission of tonal signals in a non-anechoic water tank
Authors:A E Isaev
Affiliation:(1) Institute of Ecosystem Study, National Research Council, Largo Tonolli 50, 28922 Verbania Pallanza, Italy;(2) Water Research Institute, National Research Council, Brugherio, Italy;(3) ECOCONSULT, Gavirate, Italy;(4) University of Sassari, Sassari, Italy
Abstract:An approach to estimation of the lower free-field calibration frequency as well as the required magnitude of the water tank for calibration of hydrophones in a specified frequency range from the cepstrum of a segment of the amplitude response of the electroacoustic system, projector – hydrophone – non-anechoic water tank – measurement path of standard facility, is proposed. Estimates of the lower calibration frequency of hydrophones in a water tank with minimal dimension of 6 m in the case of radiation of quadrature-added harmonic signals are presented.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号