首页 | 本学科首页   官方微博 | 高级检索  
     


Commercial spectrophotometer for particle sizing
Authors:Ferri F  Bassini A  Paganini E
Affiliation:Institute of Mathematical, Physical and Chemical Sciences, University of Milan at Como, via Lucini, 3-22100 Como, Italy.
Abstract:Particle-size distribution and the concentration of polystyrene particles suspended in water were accurately recovered from the inversion of spectral extinction data measured with a commercial spectrophotometer. The instrument was modified by placing a spatial filter in the collection optics to prevent low-angle scattered light from affecting the measurement of transmitted power. The data were inverted by use of a nonlinear iterative algorithm. When the extinction coefficient is measured in the lambda range of 0.3-1.1 mum, the particle distributions can be retrieved over a diameter range of 0.6-2.8 mum for a wide interval of sample concentrations. The average diameters are recovered with a precision of better than +/-1% and with accuracies consistent with the uncertainties by which the nominal diameters are known. The relative standard deviations of distributions corresponding to monodisperse samples are +/-5-10%, whereas the accuracy on the measured concentrations is ~5%.
Keywords:
本文献已被 PubMed 等数据库收录!