Vector characterization of photodetectors, photoreceivers, andoptical pulse sources by time-domain pulse response measurements |
| |
Authors: | Hawkins R.T. II Jones M.D. Pepper S.H. Goll J.H. Ravel M.K. |
| |
Affiliation: | Tektronix Inc., Beaverton, OR; |
| |
Abstract: | Deconvolution of measurement system effects from pulse response measurements is demonstrated to yield reproducible, accurate characterization of the impulse response (and vector frequency response) of a photodetector or photoreceiver, as well as the intensity waveform of an optical pulse. Calibration is based on a <3-ps FWHM (full width at half maximum) optical pulse and a 50-GHz 3-dB bandwidth electrical sampling system. Vector characterization of a photodetector/photoreceiver to >40 GHz and an optical pulse source to >30 GHz are demonstrated. Calibration and effects of noise are discussed |
| |
Keywords: | |
|
|