首页 | 本学科首页   官方微博 | 高级检索  
     


Thermoelectrically cooled quantum-cascade-laser-based sensor for the continuous monitoring of ambient atmospheric carbon monoxide
Authors:Kosterev Anatoliy A  Tittel Frank K  Köhler Rüdeger  Gmachl Claire  Capasso Federico  Sivco Deborah L  Cho Alfred Y  Wehe Shawn  Allen Mark G
Affiliation:Rice Quantum Institute, Rice University, Houston, Texas 77251-1892, USA. akoster@rice.edu
Abstract:We report the first application of a thermoelectrically cooled, distributed-feedback quantum-cascade laser for continuous spectroscopic monitoring of CO in ambient air at a wavelength of 4.6 microm. A noise-equivalent detection limit of 12 parts per billion was demonstrated experimentally with a 102-cm optical pathlength and a 2.5-min data acquisition time at a 10-kHz pulsed-laser repetition rate. This sensitivity corresponds to a standard error in fractional absorbance of 3 x 10(-5).
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号