首页 | 本学科首页   官方微博 | 高级检索  
     


Absolute linearity measurements on HgCdTe detectors in the infrared region
Authors:Theocharous Evangelos  Ishii Juntaro  Fox Nigel P
Affiliation:National Physical Laboratory, Queens Road, Teddington, TW11 0LW, United Kingdom. theo.theocharous@npl.co.uk
Abstract:The nonlinearity characteristics of photoconductive and photovoltaic HgCdTe detectors were experimentally investigated in the infrared wavelength region by use of the National Physical Laboratory detector linearity measurement facility. The nonlinearity of photoconductive HgCdTe detectors was shown to be a function of irradiance rather than the total radiant power incident on the detector. Photoconductive HgCdTe detectors supplied by different vendors were shown to have similar linearity characteristics for wavelengths around 10 microm. However, the nonlinearity of response of a photovoltaic HgCdTe detector was shown to be significantly lower than the corresponding value for photoconductive HgCdTe detectors at the same wavelength.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号