Study of TFTC Dynamic Character Calibration Technique |
| |
Authors: | Yong-jun Yang Jing Cai Xue-cong Zhang Jian Zhao |
| |
Affiliation: | 1. Beihang University, Beijing, China 2. Beijing Changcheng Institute of Metrology and Measurement, Beijing, China
|
| |
Abstract: | A TFTC (thin film thermocouple) is an advanced thermal sensor that is used for measuring transient temperatures. Because of the TFTC??s high-speed response, the usual calibration instrument and method cannot meet the calibration needs. This article introduces a pulse laser heating method. Dynamic character measurements and system identification are combined. The material of the TFTC is NiCr?CNiSi, and the film thickness is 2 ??m with a ceramic substrate. The transfer function is derived by means of a least-squares solution. In this method under the assumption of an ideal temperature step, there is no strict requirement on a first-order system and the limit of a pulse laser bandwidth. The calibration method introduced in this article resolves the TFTC dynamic character calibration problem. |
| |
Keywords: | |
本文献已被 SpringerLink 等数据库收录! |
|