首页 | 本学科首页   官方微博 | 高级检索  
     


Study of TFTC Dynamic Character Calibration Technique
Authors:Yong-jun Yang  Jing Cai  Xue-cong Zhang  Jian Zhao
Affiliation:1. Beihang University, Beijing, China
2. Beijing Changcheng Institute of Metrology and Measurement, Beijing, China
Abstract:A TFTC (thin film thermocouple) is an advanced thermal sensor that is used for measuring transient temperatures. Because of the TFTC??s high-speed response, the usual calibration instrument and method cannot meet the calibration needs. This article introduces a pulse laser heating method. Dynamic character measurements and system identification are combined. The material of the TFTC is NiCr?CNiSi, and the film thickness is 2 ??m with a ceramic substrate. The transfer function is derived by means of a least-squares solution. In this method under the assumption of an ideal temperature step, there is no strict requirement on a first-order system and the limit of a pulse laser bandwidth. The calibration method introduced in this article resolves the TFTC dynamic character calibration problem.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏