首页 | 本学科首页   官方微博 | 高级检索  
     


Development of an Efficient Acid Digestion Procedure Utilizing High-Pressure Asher Technique for the Determination of Iodine and Metallic Elements in Milk Powder
Authors:Kimmo Leppänen  Matti Niemelä  Paavo Perämäki
Affiliation:1. Optoelectronics and Measurement Techniques Laboratory, Department of Electrical Engineering, University of Oulu, P.O. Box 4500, 90014, Oulu, Finland
2. Department of Chemistry, University of Oulu, P.O. Box 3000, 90014, Oulu, Finland
Abstract:An effective wet digestion procedure for the determination of total iodine contents in milk powder samples was developed utilizing a high-pressure asher technique. The optimized method based on a two-stage digestion procedure. In the first stage, 500 mg samples were digested at 300 °C for 2 h using 15.2 mol L?1 HNO3 (5 mL) and 30 % H2O2 (3 mL). After the first digestion stage, digestion vessels were allowed to cool down and 1.2 mL of 20 % (w/v) Na2S2O8 solution was added as an additional oxidizing agent to the samples. After that, the vessels were closed, and they were heated at 100 °C for 30 min, resulting in clear and colorless sample solutions. Iodine concentrations in the digested samples were measured with inductively coupled plasma mass spectrometry. The accuracy of the optimized method was confirmed by analyzing milk powder reference material (BCR-151, milk powder). The obtained value for iodine (5.29?±?0.37 mg kg?1, n?=?6) was in good agreement with the certified value (5.35?±?0.14 mg kg?1). Furthermore, the results obtained for reference material showed that the developed method can be applied also for the determination of other elements, e.g., copper, iron, and lead in the digested milk powder samples.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号