Flow-compensated self-gating |
| |
Authors: | Jessica Schulz Matthias Korn Michael Deimling Wolfhard Semmler Michael Bock |
| |
Affiliation: | Siemens Medical Solutions, Erlangen, Germany. |
| |
Abstract: | OBJECTIVE: Self-gating (SG) is a method to record cardiac movement during MR imaging. It uses information from an additional short, non-spatially encoded data acquisition. This usually lengthens TE and increases the sensitivity to flow artifacts. A new flow compensation scheme optimized for self-gating sequences is introduced that has very little or no time penalty over self-gating sequences without flow compensation. MATERIALS AND METHODS: Three variants of a self-gated 2D spoiled gradient echo or fast low angle shot (FLASH) sequence were implemented: without (noFC), with a conventional, serial (cFC), and with a new, time-efficient flow compensation (sFC). In experiments on volunteers and small animals, the sequence variants were compared with regard to the SG signal and the flow artifacts in the images. RESULTS: Both cFC and sFC reduce flow artifacts in cardiac images. The SG signal of the sFC is more sensitive to physiological motion, so that a cardiac trigger can be extracted more precisely as in cFC. In a typical setting for small animal imaging, sFC technique reduces the echo/repetition time over cFC by about 23%/14%. CONCLUSION: The time-efficient sFC technique provides flow-compensated images with cardiac triggering in both volunteers and small animals. |
| |
Keywords: | Self-gating Self-navigation Flow compensation Small animal imaging |