首页 | 本学科首页   官方微博 | 高级检索  
     


Comparison of sampling techniques for in-line monitoring using Raman spectroscopy
Authors:Wikström Håkan  Lewis Ian R  Taylor Lynne S
Affiliation:Department of Industrial and Physical Pharmacy, School of Pharmacy, Purdue University, West Lafayette, Indiana 47907, USA.
Abstract:Raman spectroscopy is currently of interest as a process monitoring tool for pharmaceutical unit operations. In this study, the performance characteristics of Raman spectrometers with different sampling optics have been investigated in the context of process monitoring, with emphasis being placed on assessing homogeneity in powder blends and following changes in solid-state form during wet granulation. A novel large spot non-contact Raman sampling device was compared with a traditional small spot size non-contact sampling device and an immersion probe. The large spot non-contact optics provided significant advantages over the standard systems both as a result of the enhanced sampling volume and because of the greater robustness of the system to fluctuations in the sampling distance during the wet granulation process.
Keywords:
本文献已被 PubMed 等数据库收录!