The effect of pitch and collimation on radiation dose in spiral CT
Authors:CHENG Qi-Jun  TSANG Cheung  FENG Ding-Hua
Affiliation:1. School of Radiation Medicine and Public Health, Soochow University, Suzhou 215006, China
2. Department of Physics and Materials Science, City University of Hong Kong, Hong Kong, China
Abstract:Measurements of radiation dose to patients in spiral computed tomography (CT) were completed for various collimations,table speeds and pitch. A standard CT head dosimetry phantom and thermoluminescent dosime- ters (TLD) were used for the measurement. The effect of collimation and pitch on radiation dose was studied. The results indicated that the radiation dose at the given tube current,voltage and rotation speed was inversely propor- tional to pitch. And the increasing times of dose were as decreasing times of pitch. This regular pattern was tenable for radiation dose at both central holes and peripheral holes of the phantom at pitch = 1,>1 and <1. The collimation had no impact on the radiation dose. The results also indicated that radiation dose at central holes was nearly equal to that at peripheral holes. There was no significant difference between them statistically. The study demonstrates that the pitch in spiral CT scans is the primary parameter and has significant impact on radiation dose.
Keywords:Radiation dose  TLD  CT dosimetry head phantom  Pitch  Collimation