Multichannel radiometer calibration: a new approach |
| |
Authors: | Diaz Susana Booth Charles R Armstrong Roy Brunat Claudio Cabrera Sergio Camilion Carolina Casiccia Claudio Deferrari Guillermo Fuenzalida Humberto Lovengreen Charlotte Paladini Alejandro Pedroni Jorge Rosales Alejandro Zagarese Horacio Vernet Maria |
| |
Affiliation: | Centro Austral de Investigaciones Científicas, Consejo Nacional de Investigaciones Científicas y Técnicas, B. Houssay 200, 9410 Ushuaia, Argentina. subediaz@speedy.com.ar |
| |
Abstract: | The error in irradiance measured with Sun-calibrated multichannel radiometers may be large when the solar zenith angle (SZA) increases. This could be particularly detrimental in radiometers installed at mid and high latitudes, where SZAs at noon are larger than 50 degrees during part of the year. When a multiregressive methodology, including the total ozone column and SZA, was applied in the calculation of the calibration constant, an important improvement was observed. By combining two different equations, an improvement was obtained at almost all the SZAs in the calibration. An independent test that compared the irradiance of a multichannel instrument and a spectroradiometer installed in Ushuaia, Argentina, was used to confirm the results. |
| |