Enzyme-linked immunosorvent assay using vertical microreactor stack with microbeads |
| |
Authors: | Yoshiaki Ukita Toshifumi Asano Kuniyo Fujiwara Katsuhiro Matsui Masahiro Takeo Seiji Negoro Yuichi Utsumi |
| |
Affiliation: | 1. Laboratory of Advanced Science and Technology for Industry (LASTI), University of Hyogo, 3-1-2 Kouto, Kamigori, Ako-gun, Hyogo, 678-1205, Japan 2. Graduate School of Engineering, University of Hyogo, 2167 Shosha, Himeji, Hyogo, 671-2280, Japan
|
| |
Abstract: | The advantages of vertical microreactor stack with three-dimensional (3D) structure for immunoassay are discussed. The vertical microreactor stack uses vertical fluid flow operation with multifunctional fluid filters. The multi function of fluid filter is very effective for micromixing and passive valve operation. The mechanism of micromixing is discussed by using computational fluid dynamics (CFD), and we know that the mixing mechanism based on Coanda effect. To evaluate the micromixing performance of fluid filter, we demonstrated enzyme reaction with unique repeat mixing operation. As the results, we proved that the fluid filter has very effective mixing performance. The detection limit, which demonstrated by competition enzyme-linked immunosorvent assay (ELISA), is comparable with recommended detection limit, which suggested by Japanese ministry for the environment. |
| |
Keywords: | |
本文献已被 SpringerLink 等数据库收录! |
|